12
Alpha Scintillation Probes/Data
This 50 cm²
ZnS scintilla-
tion probe
offers excellent
alpha responses,
a low mechanical
profile and a conve-
niently angled grip. Easily-
repaired windows are provided with
protective covers for storage.
Uniformity is within ±10% over the
whole phosphor surface.
The A50L is used for general purpose
alpha monitoring. This probe can be
matched with Thermo's portable sur-
vey meters, such as the ASP-2, E-600,
RM-25, or Electra/Delta series. See
also AC-3.
• Light weight, low profile, angled grip
• Large radiation window, hex-mesh
grills, protective storage covers
• Connector, MHV
• Uniformity within ±10% over the
whole phosphor surface
• 50 cm² active area
Alpha
probe
with a
ZnS phos-
phor used for
survey and frisk-
ing in a lightweight cast
aluminum body. The radia-
tion window is 0.50 mg/cm² thick,
made of aluminized mylar. The AC-3-8
is fitted with a mesh grille. This probe
can be matched with Thermo's
portable survey meters, such as the
ASP-2, E-600, RM-25, or Electra/Delta
series.
• Operating voltage: 1,000 V nominal
• Dead time: 12 µs nominal
• Alpha efficiency (4π): 14%
239
Pu
• Operating temp: -30 ºC to 60 ºC
(-22 ºF to 140 ºF)
• Connector, CJ-1
• Size: 29.2 x 7.0 x 8.3 cm
(11.5” x 2.75” x 3.25”)
• 50 cm² window area
A50L
AC-3-7 and AC-3-8
Alpha Probe Data
Alpha efficiency is expressed as a percentage of the 2π flux of a certified reference source.
Order Code Radiation Window Efficiency Background Weight
Area
(cm²) Shape Isotope (MeV)
(% surface
emission)
s
-1
cmp
(kg) (lb)
A50L 62 rect. 230Th (4.682) >50% 0.6 1.3
AC-3-7, AC-3-8 59 rect. 0.6 1.4
AP2/4A 50 square 241Am (5.486) 35% <0.1 0.75 1.7
AP2R/4A 50 square 241Am (5.486) 25% <0.1 0.75 1.7
AP4/4A 20 disc 241Am (5.486) 35% <0.1 0.60 1.35
AP5AD 100 rect. 241Am (5.486) 35% <0.1 0.50 1.1
AP5RA 100 rect. 241Am (5.486) 23% <0.1 0.50 1.1
AP6A 600 rect. 241Am (5.486) 30% <0.1 1.55 3.5
HP-380 A 100 rect. 239Pu (5.1 MeV) 42% <0.1 <1.0 0.59 1.37