Neutron Probes
FN1 Fast Neutron Probe
For fast neutron
counting, the
phosphor uses the
(nα) reaction to
provide enhanced neutron detection
and insensitivity to gamma radiation.
The housing is made from light alloy.
Phosphor size: 1.59 x 2.54 dia. cm
(0.625” x 1” dia.)
Neutron efficiencies: 0.7 MeV, 0.01%.
3.4 MeV, 0.1%. 19 MeV, 1.6%
FN1 and TN1 are matched with
Thermo's portable survey meters,
such as the Surveyor, E-600, ASP-2,
RM-25 or Electra/Delta series.
Popular REM detector using a moderat-
ed BF
3
tube supplying excellent neutron
sensitivity and gamma rejection. This
probe can be matched with the ASP-2
and E-600. See also SWENDI-2.
Measures neutron dose rate from
thermal through fast
BF
3
tube provides high gamma
rejection
Optional mounting brackets available
22.9 cm (9”) diameter cadmium-
covered polyethylene moderating
sphere
Range: 1–60 kmRem/h
Operating Voltage: 1,600–2,000 V
Dead time: 10 µs nominal
Background: Gamma rejection
up to approximately 500 R/h
Sensitivity: ~50 cpm per mrem/h
(3000 counts per mrem)
Energy range: Thermal to
approximately 10 MeV
Operating temp:
-30 ºC to 60 ºC
(-22 ºF to 140 ºF)
Connector: MHV
Size: 25.1 x 22.9 dia. cm
(9” dia. x 9.9”)
Weight: 6.2 kg (13.75 lb)
TN1 Thermal
Neutron Probe
FN1 and TN1
weights: 9.8 oz (278 g)
Uses a proprietary phosphor, highly
efficient to slow neutrons with excel-
lent gamma rejection. MHV connector.
Phosphor size: 0.64 x 2.54 dia. cm (0.25”
x 1” dia.)
Neutron efficiencies: 0.01 eV, 60%.
0.1 eV, 30%. 1 eV, 10%.
Weight: 0.278 kg (0.61 lb)
FN1 and TN1 are matched with
Thermo's portable survey meters,
such as the Surveyor, E-600, ASP-2,
RM-25 or Electra/Delta series.
FN1 and TN1
Phosphor size:
1 in (dia.) x 0625 in
Neutron efficiencies:
0.7 MeV, 0.01%
3.4 MeV, 0.1%
19MeV, 1.6%
Phosphor size:
1 in (dia.) x 0.25 in
Neutron efficiencies:
0.01 eV, 60%
0.1 eV, 30%
1 eV, 10%
This state-of-the-art wide energy neu-
tron detector covers an extremely
wide energy range. The rugged
design and simple one step calibra-
tion makes it easy to operate and
ideal for use in various applications.
This probe can be matched with the
E-600. See also NRD.
Neutron dose equivalent rate
Improved energy response above 2
MeV
Utilizes
3
He Tube
Sensitivity: 680 cpm/mRem/h
Energy range: Thermal through 30
MeV
High voltage: 1200 to 1500 Vdc
Size: 33.7 H x 22.9 dia. cm (13.25” x
9”)
Weight: 13.38 kg (29.5 lb)
Connector: Thermo Smart Connector
for use with the E-600
(Order CA-99-XX conversion cable
for use with instruments with an
MHV connector, for example the
RMS-3)
SWENDI-2